HIGH TEMPERATURE VACUUM OR CONTROLLED ATMOSPHERE LABORATORY FURNACE
APPLICATIONS
- Brazing
- Ceramic Firing
- Degassing
|
- Heat Treating
- Melting
- Sintering
|
HOT ZONE SIZES
- 3" DIA. X 6" HIGH, THREE PHASE
- 3" DIA. X 8" HIGH, THREE PHASE
- 4" DIA. X 6" HIGH, THREE PHASE
- 4" DIA. X 8" HIGH, THREE PHASE
|
|
Temperatures up to 2500 Degrees C
Operating Pressures from 2 PSIG to 10-5 TORR
The MRF Top Loading Laboratory Furnace System comes to your lab
preprogrammed and ready to run as soon as your basic utilities have been
connnected. All water/gas plumbing and wiring are done at the factory and
your system arrives fully tested and ready to run.
The basic furnace system consists of the high vacuum chamber and heat zone
assembly, power supply, vacuum system, and control console. All these
elements are brought together and housed in an attractive package that will
complement any facility.
The furnace system can be customized to meet the most demanding
applications. Controls can vary from simple, discreet, programmable
temperature controllers to graphically orientated loop & logic computer
controls.
The basic furnace has a tungsten refractory metal hot zone. The heating
element is of a tungsten mesh design. The shielding is layered tungsten sheet.
All hot zone components are easily accessed for service. Sight ports and
thermocouple ports are provided. Additional ports can be supplied at your
request.
Power feed through ports and the vacuum port are located on the bottom and
are housed, along with gas and water plumbing, in the cabinet base.
|
|
The standard vacuum system consists of an ISO 80 diffusion pumping system
complete with a multi-coolant baffle and electropneumatic high vacuum valve.
Vacuum levels are monitored using an ionization gauge controller.
Evacuation systems are available for operation in the 10-2 Torr range and
for evacuation of the hot zone prior to backfilling with your desired inert gas.
Gas systems are available for a variety of process gases. Please consult
with the factory for special applications or requests.
The power supply is designed to accommodate a ramp rate of 50-80 deg.
C/minute, up to operating temperature. Three phase SCR control provides an
economical and efficient means of modulating power to the heating element in
a reliable and accurate fashion. Water cooled power cables supply the power
to the heating element from the main step down transformer and volt & amp
meters are supplied to visually monitor the power the furnace is using.
|
|
The MRF Commitment
Personal ServiceAccurate QuotesIntelligent
Designs
Innovative EngineeringQuality Workmanship
On-Time DeliveryDependable Performance
Technical SupportField Service
|
Return to our Home page
For truly extraordinary applications, MRF specialists can design and build
one-of-a-kind furnaces that meet unique and demanding needs of the R&D
industry. You'll appreciate the "can do" attitude at MRF.
Suncook Business Park
Rt. 28 & Lavoie Drive
Suncook, New Hampshire 03275
Voice: (603)485-2394
Fax: (603)485-2395
Email: [email protected]